Firmware Version:   

Data
Setup

Universal Non-Destructive Assay Data Acquisition Platform  UNAP
   
Status File
Management
General
Setup
Data
Setup
Time &
Communications
DCM

 

SR, Totals, & Listmode

 

SCA & Contacts

 

Trigger

SCA Setup

                 
Discriminator
(volts)
Channel
0
Channel
1
Channel
2
Channel
3
Upper
Level:
Lower
Level:

Contact Closures
 Outputs
Active Style Timer Duration
(sec.)
Output 0:
Output 1:
Output 2:
Output 3:

Inputs
Action 
Input 0: If an input contact closure is set to trigger then an associated trigger event must also be defined.
Input 1:
Input 2:
Input 3:
                 
  Single Channel Analyzer Setup

On the four Single Channel Analyze channels the upper discriminator levels must be set above the lower levels for each of the four channels

Contact Closure Outputs may be Active Open, Active Closed, or Inactive. A DAS command attempting to open or close an inactive output is logged as a configuration error. The Style may be Follow Command (no timer), Restart Timer (time is restarted if another activation while still active), and Ignore Overlap (existing timer expires as scheduled).

Contact Closure Inputs, if set to Do Trigger, will cause the configured trigger to occur. See the Trigger page to configure one or more triggers.

More...

 